LIMATHERM TTJI-6/8 Thermocouple

Temperature sensor applied in industrial pipelines, pressure tanks and different industrial devices for temperature measurement of liquids and gases. Sensing element is placed directly in the sheath.

  • Temperatura range -40÷700ºC TC
  • sensing element: J thermocouple
  • connection head type BA IP55, – 40÷100ºC
  • sheath material: acid-resistant steel 1.4541
  •  temperature transmitter can be mounted in a connection head
			int(3822)
NULL
		

Made in Poland. brand LIMATHERM, model TTJI-6/8

Fast delivery to Latvia, Estonia and Lithuania. International delivery to Europe, Asia, Africa and other continents available on request
Our warehouse is located in Riga, Latvia.

price by request

Available

Available

Advantages

Official distributor in Latvia

Time-tested American, European and Asian brands of control and measurent instruments

Advantages

Special price list for organisations

We offer a significant discount for trade organisations and manufacturing companies

Advantages

Worldwide shipping on request

We take care of all logistics and customs clearance. Fast delivery to Baltic countries

Advantages

After-sales service and support

Warranty service and repair, expert assistance in the operation of the instruments

Technical specifications of TTJI-6/8

Measuring range* / sensing element (-40 ÷ 700) ºC J class 2
Measuring insert
  • non-replaceable
Sheath
  • material: steel 1.4541
  • sheath length L [mm]:
    standard – 115, 175, 245, 375, 525 for ø6, Lmax. [mm]: 1500
    standard – 495, 705, 995, 1395, 1995 for ø8, Lmax. [mm]: 2000
Connection head BA, IP55, (-40 ÷ 100) ºC
Options
  • local display built into the DANAWwin head connection p. 220
  • measuring junction types p. 13
  • connection heads: p. 217÷218 (eg. stainless BEG; aluminium NA,
    IP65; NA with snap lock)
  • TC: class 1
Additional accessories
  •  temperature transmitters p. 225÷241
  • process protection tubes and thermowells p. 200÷214
  • compensation cables p. 197
  • mounting brackets: UG-1, UG-3, UG-8, UZK-1 p. 215÷216
Load More Items

Time response T05/T09

Sensor type ø8
J insulated junction ≤22/≤62
J grounded junction ≤3/≤8
Load More Items

Resistor tolerance acc. to PN-EN 60751

Wire wound resistor
Range [°C] Tolerance [°C]
AA (-50÷250) ±(0,1+0,0017∙|t|)
A (-100÷450) ±(0,15+0,002∙|t|)
B (-196÷600) ±(0,3+0,005∙|t|)
Load More Items

Tolerance for thermocouples class acc to. PN-EN 60584

Class 1 Class 2
Range [°C] Tolerance [°C] Range [°C] Tolerance [°C]
J
Fe-CuNi
 (-40÷375)
(375÷750)
±1,5
±0,004 |t|
 (-40÷333)
(333÷750)
±2,5
±0,0075 |t|
Load More Items

How to buy TTJI-6/8

Ordering example: TOPI-6115B2

User Guide and Instruction manual for LIMATHERM TTJI-6/8

You can request the necessary certificates, documentation, instructions and user guide for LIMATHERM TTJI-6/8 using the special form.

You can buy LIMATHERM TTJI-6/8 in Latvia at a favourable price in «OLIL SIA» by leaving a request on the website, by phone +371 672 70580 or by email . LIMATHERM TTJI-6/8 price are subject to change depending on exchange rates. We deliver instruments by transport companies in Latvia, Estonia and Lithuania. Worldwide shipping to Europe, Asia, Africa, North and South America available on request.