MSA PrimaX IR+ Gas Detector

  • Measuring Range 0 – 100% LEL
  • Outputs
3-Wire, 4-20 mA current source, HART 7 protocol compatible
3 relays: 2 alarm, 1 fault (2A @ 30 VDC)
  • Protection Rating Dual IP66 and IP67
  • Temperature Range -40°C to 75°C (-40°F to 167°F)
			int(17085)
NULL
		

Made in USA. brand MSA Safety, model PrimaX IR+

Fast delivery to Latvia, Estonia and Lithuania. International delivery to Europe, Asia, Africa and other continents available on request
Our warehouse is located in Riga, Latvia.

price by request

Available

Available

Advantages

Official distributor in Latvia

Time-tested American, European and Asian brands of control and measurent instruments

Advantages

Special price list for organisations

We offer a significant discount for trade organisations and manufacturing companies

Advantages

Worldwide shipping on request

We take care of all logistics and customs clearance. Fast delivery to Baltic countries

Advantages

After-sales service and support

Warranty service and repair, expert assistance in the operation of the instruments

Descriptions of MSA PrimaX IR+

  • Patented dual source design – A redundant Infrared (IR) source provides reliability and uninterrupted monitoring should a source failure occur. In addition, optics are optimized for maximum signal, resulting in a product of extraordinary stability.
  • Flexible output options  – The unit is equipped with LED display, status indicators and relay outputs providing the users with flexible combustible gas detection monitoring using 4-20 mA output and HART enabled diagnostics.
  • Remote sensing capability – The sensor can be remotely mounted up to 30 meters/100 feet from the display unit using Aluminum or Stainless Steel optional junction boxes.
  • Thoroughly tested and performance approved – SIL2 suitable for use in critical Safety Instrumented Systems and IEC/EN 60079-29-1 performance approved for various combustible gases.

Technical specifications of MSA PrimaX IR+

General specifications
Measuring Range  0 – 100% LEL
Gas Types   Hydrocarbon gases and vapours;
The PrimaX IR+ Detector is available with factory
Methane and Propane calibrations, the instrument
may also be calibrated to following performance
approved gases:
Methane, Propane, Toluene, n-Butane, n-Pentane,
n-Hexane, Propylene, Ethane, Propylene Oxide,
Acetone, Cyclopentane, Ethyl Acetate, IsoButane,
Ethylene. Most of other flammable hydrocarbon
vapours are also detectable.
Warranty  3 years full instrument

10 years IR source

Signal Outputs  3-Wire, 4-20 mA current source, HART 7 protocol
compatible
3 relays: 2 alarm, 1 fault (2A@30 VDC)
Output Levels
(default values)
Normal for 0 – 100% LEL 4.0 – 20.0 mA
Safety Critical Fault 0.0 mA
*Fault  2.0 mA
*Obscuration or dirty optics 2.5 mA
*Power up or calibration 3.0 mA
*Cleaning mode 3.5 mA
Overrange 20.0 – 20.5 mA
(*) user configurable via the HART interface
Current Draw  175 mA RMS average @ 24 VDC

250 mA Max in-rush @ 24 VDC including sensor

Power Input  20-30 VDC, 6 Watts max
Wiring requirement Connection terminals: 3-Wires, 2.5 mm2 max
Performance specifications
Response Time
Methane/Propane
T50 < 5 sec, T90 < 7 sec (without environmental guard)
T50 < 7 sec, T90 < 15 sec (with environmental guard)
Stability  +/- 2% Full Scale/Year
Repeatability  +/- 1% Full Scale
Mechanical specifications
Physical  320-W x 152-L x 115-H mm (12.6 x 6.0 x 4.5 inches)

Weight with display and mounting bracket:
SS: 3.8 kg (8.4 lb) & AL: 2.8 kg (6.2 lb)

Material Type  Display housing: Aluminum or 316 Stainless Steel with 3 each M25 entries (316 Stainless Steel stopping plug provided)
Sensor assembly: 316 Stainless Steel
Mounting bracket
(included)
316 Stainless Steel, with 2” pipe mount U-Bolts
Environmental specifications
Temperature range -40 to 75°C (-40 to 167°F)
Humidity  15-95% RH, non-condensing
Ingress Protection  Dual IP66 and IP67 rated
Approvals
General
IECEx, ATEX, UKCA, CE Marking, SIL 2 suitable
Approval Ratings
European Approvals
ATEX Directive 2014/34/EU, EMC Directive
2014/30/EU
Gas II 2 G Ex db IIC T4 Gb
Dust II 2 D Ex tb IIIC T130°C Db
-40°C ≤ Ta ≤ +75°C
EN 60079-0:2018, EN 60079-1:2014, EN 60079-31:2014, EN 60079-29-1:2016, EN 50271:2018
IEC Approvals
Gas Ex db IIC T4 Gb
Dust Ex tb IIIC T130°C Db
-40°C ≤ Ta ≤ +75°C
IEC 60079-0:2017, IEC 60079-1:2014, IEC 60079-31:2013, IEC 60079-29-1:2016
Safety Integrity Level 
Suitable for SIL2 (HFT=0), SIL3 (HFT=1) according to IEC 61508
Load More Items

 

User Guide and Instruction manual for MSA Safety PrimaX IR+

You can request the necessary certificates, documentation, instructions and user guide for MSA Safety PrimaX IR+ using the special form.

You can buy MSA Safety PrimaX IR+ in Latvia at a favourable price in «OLIL SIA» by leaving a request on the website, by phone +371 672 70580 or by email . MSA Safety PrimaX IR+ price are subject to change depending on exchange rates. We deliver instruments by transport companies in Latvia, Estonia and Lithuania. Worldwide shipping to Europe, Asia, Africa, North and South America available on request.