+371 672 70580

Mon-Fri: 9:00 - 17:00 LV

Email:

TSI FHM10 Fume Hood Monitor

Fume hoods are a primary source of protection in laboratories.

			int(17740)
NULL
		

Made in USA. brand TSI, model FHM10

Fast delivery to Latvia, Estonia and Lithuania. International delivery to Europe, Asia, Africa and other continents available on request
Our warehouse is located in Riga, Latvia.

price by request

Available

Available

Advantages

Official distributor in Latvia

Time-tested American, European and Asian brands of control and measurent instruments

Advantages

Special price list for organisations

We offer a significant discount for trade organisations and manufacturing companies

Advantages

Worldwide shipping on request

We take care of all logistics and customs clearance. Fast delivery to Baltic countries

Advantages

After-sales service and support

Warranty service and repair, expert assistance in the operation of the instruments

Descriptions of TSI FHM10

Face velocity measurements are often used to gauge the performance of a fume hood’s ability to contain and exhaust hazardous chemicals. By accurately measuring face velocity and alarming on unsafe conditions, TSI’s FHM10 Fume Hood Monitors provide a high level of fume hood safety. And, the FHM10 helps users comply with standards and regulations set forth by OSHA, NFPA, ANSI Z9.5, and SEFA, calling for monitors to be installed on fume hoods.
Applications
▪ Research Laboratories
▪ Life Science and Pharmaceutical
▪ Universities and Academic
▪ Vivariums
▪ Healthcare Facilities

Features and Benefits
▪ Accurately measures average fume hood face velocity
▪ Temperature compensated sensor for accurate measurement as temperatures change in laboratories
▪ Visual, audible, and remote alarms warn users of unsafe conditions
▪ Analog output provides additional information
▪ Assists in managing risk by communicating fume hood status information to Building Management System (BMS)
▪ Seamless integration to BMS via BACnet, LonWorks, or Modbus
▪ Easy installation and calibration
▪ Easy configuration using keypad
▪ Large display provides detailed fume hood information
▪ Surface or flush mount options available
▪ Optional Flow Monitor available

Technical specifications of TSI FHM10

  • Display Range
    0 to 1,000 fpm (0 to 5.08 m/s)
    0 to 10,000 cfm (0 to 4,720 l/s, 0 to 16,990 m3/hr)
  • Low Alarm Range
    5 to 980 fpm (0.03 to 4.98 m/s)
    0 to 10,000 cfm (0 to 4,720 l/s, 0 to 16,990 m3/hr)
  • High Alarm Range
    80 to 1,000 fpm (0.41 to 5.08 m/s)
    0 to 10,000 cfm (0 to 4,720 l/s, 0 to 16,990 m3/hr)
  • Analog Outputs
    0-10 VDC or 4-20 mA
    Represents Face Velocity, or Flow Rate
  • Alarm Contact Outputs – SPST, 2A @ 30 VDC Nominal
  • Contact Inputs – Night Setback, Flow
  • Communcation Options – Modbus, N2, BACnet MS/TP, LonWorks
  • Input Power  – 24 VAC, 50/60 Hz or 15-40 VDC 5, Watt Maximum
  • Operating Temperature – 32 to 120° F (0 to 48.9° C)
  • Size (H x W x D) – 6.67” x 2.92” x 1.25” (16.9 cm x 7.4 cm x 3.2cm)
  • Weight – 0.5 lb (225 g)
  • Optional Accessories
    800920 Slimline Monitor
    800926 Flush Mounting Bracket

User Guide and Instruction manual for TSI FHM10

You can request the necessary certificates, documentation, instructions and user guide for TSI FHM10 using the special form.

You can buy TSI FHM10 in Latvia at a favourable price in «OLIL SIA» by leaving a request on the website, by phone +371 672 70580 or by email . TSI FHM10 price are subject to change depending on exchange rates. We deliver instruments by transport companies in Latvia, Estonia and Lithuania. Worldwide shipping to Europe, Asia, Africa, North and South America available on request.