+371 672 70580

Mon-Fri: 9:00 - 17:00 LV

Email:

DWYER Magnesense AT-MSX atex/iecex approved differential pressure transmitter

The Series AT-MSX ATEX/IECEx Approved Magnesense® Differential Pressure Transmitter is an extremely versatile transmitter for monitoring pressure and air velocity in hazardous areas. This transmitter is loaded with features such as: field selectable English or metric ranges, field upgradeable LCD display, adjustable dampening of output signal and the ability to select a square root output for use with pitot tubes and other similar flow sensors. Along with these features, the piezo sensing technology provides long-term performance and enables the Magnesense® transmitter to be the solution for a myriad of pressure and flow applications. Flameproof enclosures are available in aluminum and can include a glass window for viewing process on the LCD.

			int(2746)
NULL
		

Made in USA. brand DWYER, model Magnesense AT-MSX

Fast delivery to Latvia, Estonia and Lithuania. International delivery to Europe, Asia, Africa and other continents available on request
Our warehouse is located in Riga, Latvia.

price by request

Available

Available

Founded in 1931, DWYER manufactures equipment for measurement, conversion and control – pressure, temperature, level, flow.

Advantages

Official distributor in Latvia

Time-tested American, European and Asian brands of control and measurent instruments

Advantages

Special price list for organisations

We offer a significant discount for trade organisations and manufacturing companies

Advantages

Worldwide shipping on request

We take care of all logistics and customs clearance. Fast delivery to Baltic countries

Advantages

After-sales service and support

Warranty service and repair, expert assistance in the operation of the instruments

Description of DWYER Magnesense AT-MSX

PRODUCT APPLICATIONS

  • Monitor pressures in ducts, rooms, or total building pressure
  • Filter monitoring
  • Local indication of clean room pressures with process signal sent to control room
  • Hazardous area pressure measurement and transmitter

FEATURES

  • All the capabilities and value of the MSX in an ATEX/IECEx approved enclosure
  • Long service life and minimized downtime due to durable, rugged housing and high-quality components
  • High impact strength and high temperature rated for applications where hazardous environments exist

Technical specifications of DWYER Magnesense AT-MSX

  • Service: Air and non-combustible, compatible gases.
  • Wetted Materials: Consult Factory.
  • Accuracy: ±1% FSO.
  • Stability: ±1% FSO/year.
  • Temperature Limits: -4 to 158°F (-20 to 70°C).
  • Pressure Limits: See chart on catalog page.
  • Power Requirements: 10-36 VDC (2-wire), 17-36 VDC or isolated 21.6-33 VAC (3-wire).
  • Output Signal: 4-20 mA (2-wire); 0-10 V or 0-5 V selectable (3-wire).
  • Response Time: Instantaneous (default) or 3 s (selectable) without considering delay caused by Exd flame arrestors (up to 30 s from FS to zero depending model and ranges with STD pressure ports) about 15 s with LD pressure ports on very low ranges.
  • Zero and Span Adjustments: Digital push-button.
  • Loop Resistance: Current output: 0 to 1250 Ω max.; Voltage output: min. load resistance 1 kΩ.
  • Current Consumption: 21 mA max continuous.
  • Display: 4 digit LCD.
  • Electrical Wiring: 4-wire removable European style terminal block for 16 to 26 AWG.
  • Mounting Orientation: Pressure sensor measurement unaffected by orientation.
  • Enclosure Rating: IP66.
  • Housing Material: Aluminum.
  • Finishing: Texture epoxy coat RAL7038.
  • Process Connections: 1/8″ NPT female brass (SS optional).
  • Electrical Connections: Two 1/2″ NPT female. Cable gland not included.
  • Weight: 12.3 lb (5.6 kg).
  • ATEX Certificate:
  • INERIS 21ATEX0033X.
  • IECEx Certificate:
  • IECEx INE 21.0064X.
  • Compliance: ATEX: CE 0080 II 2G Ex db IIC T5, T6 Gb -60°C≤Ta≤+50°C (T6) -60°C≤Ta≤+60°C (T5); II 2D Ex tb IIIC T75°C Db IECEx: Ex db IIC T5, T6 Gb -60°C≤Ta≤+50°C (T6) -60°C≤Ta≤+60°C (T5) Ex tb IIIC T75°C Db.

Model Chart DWYER Magnesense AT-MSX

AT-MSX MAGNESENSE PR XMTR W/ ATEX ENCLOSURE
     
     AT-102NA-MSX-XX-X-X-XXXX  
     
          1ST SYMBOL (1ST X) DIRECTION
             1    UNI-DIRECTIONAL
             2    BI-DIRECTIONAL
             
          2ND SYMBOL (2ND X) PRESSURE RANGE
             0    0.5 INWC, 125 PA, 12 MMWC
             1    1 INWC, 250 PA, 25 MMWC
             2    5 INWC, 1250 PA, 125MMWC
             3    28 INWC, 7000 PA, 700 MMWC
             
          3RD SYMBOL (3RD X) UNITS OF MEASURE
             IN   INCHES OF WATER COLUMN
             PA   PASCAL
             MM   MILLIMETERS OF WATER COLUMN

          4TH SYMBOL (4TH X) DISPLAY	
             LCD  WITH LCD
  
          5TH SYMBOL (5TH X) COVER
             B    BLIND
             W    GLASS WINDOW

          6TH SYMBOL (6TH X) PRESSURE PORT/VALVE MATERIAL
             1    BRASS
             2    STAINLESS STEEL
  
          7TH SYMBOL (7TH X) PRESSURE PORT/VENTING VALVE CONFIGURATION
             VS0  STD PRESSURE PORT/NO VALVE
             VS1  STD PRESSURE PORT/STD VALVE
             VS2  STD PRESSURE PORT/LD VALVE
             
          8TH SYMBOL (8TH X) CABLE ENTRY
             12   1/2” NPT ANSI/ASME B1.20.1

User Guide and Instruction manual for DWYER Magnesense AT-MSX

You can request the necessary certificates, documentation, instructions and user guide for DWYER Magnesense AT-MSX using the special form.

You can buy DWYER Magnesense AT-MSX in Latvia at a favourable price in «OLIL SIA» by leaving a request on the website, by phone +371 672 70580 or by email . DWYER Magnesense AT-MSX price are subject to change depending on exchange rates. We deliver instruments by transport companies in Latvia, Estonia and Lithuania. Worldwide shipping to Europe, Asia, Africa, North and South America available on request.