TSI 4070 Certifier Flow Analyzer Ventilator Test System

Low-cost flow analyzer system capable of testing multiple parameters of ventilator performance

			int(17810)
NULL
		

Made in USA. brand TSI, model 4070

Fast delivery to Latvia, Estonia and Lithuania. International delivery to Europe, Asia, Africa and other continents available on request
Our warehouse is located in Riga, Latvia.

price by request

Available

Available

Advantages

Official distributor in Latvia

Time-tested American, European and Asian brands of control and measurent instruments

Advantages

Special price list for organisations

We offer a significant discount for trade organisations and manufacturing companies

Advantages

Worldwide shipping on request

We take care of all logistics and customs clearance. Fast delivery to Baltic countries

Advantages

After-sales service and support

Warranty service and repair, expert assistance in the operation of the instruments

Descriptions of TSI 4070

The 4070 Certifier™ FA System is made up of a 4076 Certifier FA High-Flow Module Kit and an 4078 Certifier FA Interface Display Module Kit.

Features and Benefits
Simple, easy to read, user interface
Backlit LCD display
Battery operated (AA batteries)
Volumes and flow in BTPS, ATP, or STP
Compact design, entire kit weighs less than 3 lbs (1.4 Kg)

Technical specifications of TSI 4070

Physical

Dimensions Controller module: 13.2 cm x 7 cm x 3.3 cm (5.2 in. x 2.8 in. x 1.3 in.).

High Flow module: 18 cm x 6.6 cm x 4 cm (7.1 in. x 2.6 in. x 1.6 in.).

Low Flow module: 12.7 cm x 5.1 cm x 2.8 cm (5.0 in. x 2.0 in. x 1.1 in.).

Flow connectors High Flow module:

•         flow inlet: 22-mm female ISO taper.

•         flow outlet: 22-mm male ISO taper.

Low Flow module:

•         flow inlet: 0.25-in. barb.

•         flow outlet: 0.25-in. barb.

Weight Approximately 1.36 kg (3 lb) for the standard kits.
Load More Items

 Environmental

Temperature Operating: 5 to 40 ℃ (41 to 104°F). 15 to 80% relative humidity from 5 to 31 °C decreasing linearly to 15 to 50% relative humidity at 40 °C. Storage: -40 to 70°C (-40 to 158 °F) at 10 to 95% relative humidity.
Atmospheric Pressure Operating: 57.1 to 106 kPa (8.28 to 15.37 psia).

Storage: 50 to 106 kPa (7.25 to 15.37 psia).

Conditions Indoor Use

Operating Altitude up to 4000 m (13,000 ft) Pollution degree I or II

Load More Items

Power

Battery life 15 to 20 hours.
Battery type Four commercially available AA batteries.
Load More Items

Test Measurements (see notes at end of section)

Measurement High Flow Module Low Flow Module
Flow and Peak Flow  
Range 0 to 300.0 SLPM. 0.01 to 15.00 SLPM.
Accuracy Air and oxygen: ± 2% of reading or ± 0.075 SLPM, whichever is greater.

Air/oxygen mixtures: ± 4% of reading or ± 0.1 SLPM, whichever is greater.

Air and oxygen: ±2% of reading or ± 0.010 SLPM, whichever is greater.

Nitrous oxide: ± 4% of reading or ± 0.025 SLPM, whichever is greater.

Measurement High Flow Module Low Flow Module
Pressure drop Maximum between inlet and outlet ports at 101.3 kPa (14.7 psia) including filter: 0.50 cmH2O at 20 SLPM, 1.50 cmH2O at 50 SLPM, 5.0 cmH2O at 100 SLPM, 30.0 cmH2O at 300 SLPM. Maximum between inlet and outlet ports at 101.3 kPa (14.7 psia) including filter: 4 cmH2O at 2 SLPM, 16 cmH2O at 5 SLPM, 45 cmH2O at 10 SLPM, 85 cmH2O at 15 SLPM.
Volume
Range 0.01 to 10.0 L STP. 0 to 9.999 L STP.
Accuracy Air and oxygen: ± 2% of reading plus 0.020 L STP

Air/oxygen mixtures: ±4% of reading plus 0.020 L STP.

Air and oxygen: ±2% of reading or ± 0.010 L STP, whichever is greater.

Nitrous oxide: ±4% of reading or ±0.010 L STP, whichever is greater.

Minute Volume
Range 0 to 99.00 L STP. 0 to 9.999 L STP.
Accuracy  7% of reading.  7% of reading.
Stacked Volume
Range 0 to 99.00 L STP. 0 to 9.999 L STP.
Accuracy Air and oxygen: ± 2% of reading plus 0.020 L STP

Air/oxygen mixtures: ± 4% of reading plus 0.020 L STP.

Air and oxygen: ±2% of reading or ±0.010 L STP, whichever is greater.

Nitrous oxide: ± 4% of reading or ± 0.010 L STP, whichever is greater.

Inspiratory Time
Range 0.25 to 60.00 seconds. 0.25 to 60.00 seconds.
Accuracy ±0.01 seconds. ±0.01 seconds.
I:E Ratio
Range 1:100.0 to 100.0:1. 1:15.0 to 15.0:1.
Accuracy ± 5% of reading. ± 5% of reading.
Respiratory Rate
Range 0.5 to 120.0 breaths per minute. 0.5 to 120.0 breaths per minute.
Accuracy ± 5% of reading. ± 5% of reading.
Low Pressure
Range -25.0 to 150.0 cmH2O.

(-18.4 to 110 mmHg)

Not applicable.
Accuracy ±0.75% of reading or

±0.20 cmH2O (0.15 mmHg), whichever is greater.

Not applicable.
Load More Items
Measurement High Flow Module Low Flow Module
Peak & PEEP Low Pressure
Range 0 to 150.0 cmH2O.

(0 to 110 mmHg)

Not applicable.
Accuracy ± 0.75% of reading or ± 0.20 cmH2O (0.15 mmHg), whichever is greater. Not applicable.
Barometric Pressure
Range 375 to 1500 mmHg. 375 to 1500 mmHg.
Accuracy ± 8 mmHg.  8 mmHg.
Oxygen Concentration
Range 0 to 100.0% O2. Not applicable.
Accuracy ± 2% O2 at daily calibration conditions. Not applicable.
NOTES:

1.       Standard conditions are defined as 21.1 °C (70 °F) and 101.3 kPa (14.7 psia).

2.       Flow and volume accuracy is applicable in SLPM or STP mode only.

3.       The temperature of the gas and the ambient air must be within ±10 °C (± 18 °F) of each other and the gas must be less than 30% relative humidity at 21. 1 ° C (70 ° F).

4.       Flow and volume accuracy de-rating: ±0.075% of reading per 1 ° C (1.8 ° F) away from 21.1 ° C (70 ° F); ± 0.015% of reading per 1.03 kPa (0.15 psia) above 101.3 kPa (14.7 psia); ±0.022% of reading per 1.03 kPa (0.15 psia) below 101.3 kPa (14.7 psia); ± 0.07% of reading per 1% relative humidity above 30% relative humidity.

Load More Items

 Calibration Requirements

Flow modules Factory calibration every year for normal use under normal conditions using the filter(s) provided.
Controller module No calibration required.
Oxygen sensor Daily, following sensor replacement or as required.
Load More Items

Compliance and Approvals

Complies with these standards: CAN/CSA-C22.2 No. 1010. 1-92: Canadian Standard for the Safety of Electrical Equipment for Measurement, Control and Laboratory Use, Part 1.
  EN 55011 (1991) Class B, CISPR 11 (1990) Class B, FCC (CFR 47, Part 15) Class B: Emissions, Radiated and Conducted.
  EMC Directive 89/336/ECC, EN 61326-1 (1997 plus Amendment A1 1998), IEC 1000-4-2 (1995), EN 61000-4-2, IEC 1000-4-3 (1995), EN 61000-4-3: Immunity.
Load More Items

User Guide and Instruction manual for TSI 4070

You can request the necessary certificates, documentation, instructions and user guide for TSI 4070 using the special form.

You can buy TSI 4070 in Latvia at a favourable price in «OLIL SIA» by leaving a request on the website, by phone +371 672 70580 or by email . TSI 4070 price are subject to change depending on exchange rates. We deliver instruments by transport companies in Latvia, Estonia and Lithuania. Worldwide shipping to Europe, Asia, Africa, North and South America available on request.