+371 672 70580

Mon-Fri: 9:00 - 17:00 LV

Email:

LIMATHERM W..P-EXII RTD

Mineral insulated metal sheathed insert is used as a replaceable element of temperature sensors. Insert design enables its installation in the connection head and pressing insert tip to the bottom of the sensor thermowell.

  • for intrinsically safe (Exi) temperature sensors produced by Limatherm Sensor acc. to ATEX
  • sheath material: acid-resistant steel 1.4541
			int(8750)
NULL
		

Made in Poland. brand LIMATHERM, model W..P-EXII

Fast delivery to Latvia, Estonia and Lithuania. International delivery to Europe, Asia, Africa and other continents available on request
Our warehouse is located in Riga, Latvia.

price by request

Available

Available

Advantages

Official distributor in Latvia

Time-tested American, European and Asian brands of control and measurent instruments

Advantages

Special price list for organisations

We offer a significant discount for trade organisations and manufacturing companies

Advantages

Worldwide shipping on request

We take care of all logistics and customs clearance. Fast delivery to Baltic countries

Advantages

After-sales service and support

Warranty service and repair, expert assistance in the operation of the instruments

Technical specifications of W..P-EXII

Measuring range / sensing element:

  • (-200÷600) ºC / Pt100 / class B

Version I, pipe execution:

  • material: steel 1.4541, sheath diameter [mm]: 6, 8
  • operating temperature: max. 600°C
  • standard sheath lengths Lw[mm]:
    • 115, 175, 245, 375, 525[mm] forø6,
    • 495, 705, 995, 1395, 1995[mm] forø8, or other*
  • Lmax. [mm]= 1500 forø6
  • Lmax. [mm]= 2000 forø8

Version II, mineral insulated execution: max. operating temperature:

  • RTD Pt100 materia l1.4571
  • length L[mm]: acc. to requirements
ø3 ø4,5 ø6 ø8
450°C 550°C 700°C
900°C 1000°C 1200°C 1200°C
400°C 600°C
Load More Items

Options:

  • Temperature transmitter application: ATEX approved temperature transmitter with standard 4÷20mA output signal and with the HART communication protocol can be mounted in the connection head. Transmitter installation is carried out directly on the measuring insert (in place of a terminal block) or in the high cover connection head (solution used to enable installation of two transmitters).
  • Non-standard design: Immersion length and the measuring insert parameters can be customized per client request.

Response time to temperature change:

Thermowell diameter
[mm
Response time [s]
ø6 t0,5= 12
t0,9= 55
ø8 t0,5= 20
t0,9= 85
Load More Items

Tolerance for classes of sensors with resistors Pt acc. to PN-EN 60751

Sensor
classes
Range of application [°C] Formula for calculating
acceptable deviations [°C]
AA -50÷250 T = ±(0,10 + 0,0017 |t|)
A -50÷450 T = ±(0,15 + 0,002 |t|)
B -196÷600 T = ±(0,3 + 0,005 |t|)
Load More Items

How to buy W..P-EXII

Ordering example:  APW2K/3–5000–1–SO–FlexTop2211-Ex– (0÷800)°C–Exi

User Guide and Instruction manual for LIMATHERM W..P-EXII

You can request the necessary certificates, documentation, instructions and user guide for LIMATHERM W..P-EXII using the special form.

You can buy LIMATHERM W..P-EXII in Latvia at a favourable price in «OLIL SIA» by leaving a request on the website, by phone +371 672 70580 or by email . LIMATHERM W..P-EXII price are subject to change depending on exchange rates. We deliver instruments by transport companies in Latvia, Estonia and Lithuania. Worldwide shipping to Europe, Asia, Africa, North and South America available on request.