Made in Germany. brand SIKA, model TP37450E.2
Fast delivery to Latvia, Estonia and Lithuania. International delivery to Europe, Asia, Africa and other continents available on request
Our warehouse is located in Riga, Latvia.
price by request
Available
Official distributor in Latvia
Time-tested American, European and Asian brands of control and measurent instruments
Special price list for organisations
We offer a significant discount for trade organisations and manufacturing companies
Worldwide shipping on request
We take care of all logistics and customs clearance. Fast delivery to Baltic countries
After-sales service and support
Warranty service and repair, expert assistance in the operation of the instruments
Whether for on-site use in the workshop, in the measuring and test room or laboratory, SIKA temperature calibrators can be used everywhere. A wide range of industries and various application areas are covered.
• assembly, commissioning
• manufacturing, production
• maintenance, service
• quality assurance, test equipment monitoring
• repair
Dry block function
The dry block function was developed to enable easy calibration of the temperature in lab and field applications. The optimum thermal coupling from block to DUT is achieved by the correct calibration insert. Ideally, the inner diameter of the bore hole in the calibration insert is 0.5 mm larger than the outer diameter of the DUT. With the help of the calibration insert almost all straight temperature sensors with different lengths and diameters can be calibrated. The dry block covers the entire temperature range without changing the calibration medium. The user does not have to take viscosities, flash points or outgassing into consideration.
Dark-radiation function
A patented infrared calibration insert is used to calibrate IR pyrometers or thermal imaging cameras. The special surface structure and the asymmetrical shapes form a “black cavity radiator” with an emission factor of 0.9994 and prevent the reflection of interfering radiation and emit the desired temperature in an ideal form. The pyrometer is simply held at the prescribed distance above the measuring port and thus forms the desired measuring spot for the calibration to be carried out on the ground. The direct mounting of a tripod is possible.
Surface temperature function
Surface temperature sensors are calibrated by means of special calibration inserts which are placed vertically with the required contact pressure. The best possible temperature reference point on the surface of the calibration insert is found by switching the calibrator control to the external reference temperature sensor. This is located directly under the front face of the calibration insert. The calibration insert is designed in such a way that the best temperature homogeneity is achieved in the middle of the front face. A good thermal contact is enabled by the special surface structure of the front face. The use of heat-conducting paste or other heat transfer media is not necessary.
Accessories
The extensive range of accessories for the TP Premium series enables time-saving calibration setups. For example, DUT can be identified via barcode scanner or an ongoing or already completed calibration process can be displayed via WLAN connection on a PC or smartphone anywhere in the world. Thanks to the possibility of connecting an external measuring bridge directly to the calibrators, many DUT can be calibrated at the same time – including automatic generation of individual SIKA works calibration certificates.
TP 37450E.2 / TP 37450E.2i | ||||
Temperature range | Room temperature…450 °C | |||
Dimension of the calibration insert | Ø 60 x 150 mm (calibration insert easily exchangeable) | |||
Dry block Air Shield Insert | External reference temperature sensor | |||
Display accuracy | ±0.2 °C | |||
Temperature stability | ±0.010 °C | |||
Temperature distribution
g Axial g Radial |
±0.100 °C
±0.015 °C |
|||
Influence of load | ±0.010 °C | |||
Dry block | External reference temperature sensor Internal reference temperature sensor | |||
Display accuracy | ±0.25 °C | ±0.45 °F | ||
Temperature stability | ±0.010 °C | ±0.018 °F | ||
Temperature distribution
g Axial g Radial |
±0.300 °C
±0.060 °C |
±0.54 °F
±1.08 °F |
||
Influence of load | ±0.015 °C | ±0.027 °F | ||
Infrared calibration | External reference temperature sensor Internal reference temperature sensor | |||
Display accuracy | ±0.5 °C | ±0.9 °F | ||
Temperature stability | ±0.030 °C | ±0.054 °F | ||
Emission factor | 0,9994 | |||
Surface calibration | External reference temperature sensor | |||
Display accuracy | ±1 °C | |||
Temperature stability | ±0.250 °C | |||
General data | ||||
Stabilisation time
(with external reference temperature sensor) |
||||
g to ±0.05°C
g to ±0.005°C |
g to ±0.09 °F
g to ±0.009 °F |
From 1 min
From 5 min |
||
Heating time | ||||
g 20 °C…440 °C | g 68…824 °F | 17 min | ||
Cooling time | ||||
g 450 °C…30 °C | g 824…86 °F | 36 min | ||
Resolution of the temperature display | 0.001 °C | |||
Hysteresis | ±0.010 °C | |||
Temperature units | °C / °F / K (selectable) | |||
Reference temperature sensor | Internal / external (selectable) | |||
Interfaces | Ethernet, 3 x USB | |||
Connectivity | OPC UA, HTTP. Details and further possibilities on request. | |||
Dimensions | ||||
g Width
g Height g Depth |
210 mm
330 mm + 50 mm (Handle) 300 mm |
|||
Weight | 11 kg | |||
Power supply | 110…115 V 60 Hz / 230 V 50 Hz
Protective conductor (PE) needed |
|||
Power consumption | Approx. 1000 W | |||
Adjustable temperature range | 0…450 °C | |||
Display | Brilliant color touchscreen (7 inches), multi panel safety glass |
You can request the necessary certificates, documentation, instructions and user guide for SIKA TP37450E.2 using the special form.
You can buy SIKA TP37450E.2 in Latvia at a favourable price in «OLIL SIA» by leaving a request on the website, by phone +371 672 70580 or by email olil@olil.lv. SIKA TP37450E.2 price are subject to change depending on exchange rates. We deliver instruments by transport companies in Latvia, Estonia and Lithuania. Worldwide shipping to Europe, Asia, Africa, North and South America available on request.