+371 672 70580

Mon-Fri: 9:00 - 17:00 LV

Email:

NAGMAN METCAL 40 Plus Metrological Dry Block Sub-Zero Temperature Calibrator

  • Advanced – Micro-controller based Technology
  • Heavy Duty Portable / Bench-top Model
  • High Accuracy, Resolution & Stability
  • Large Graphic Display (Multi-information Display)
  • Measurement Capability for TC, RTD, mV / mA / V & Resistance Signals
  • Touch Membrane Keypad
  • Switch Test Contact Measurement
  • Multi-hole Thermowell Inserts
  • Computer Interface & Calibration Software
  • Count down timer (Shows when unit is stable)
  • Stability Indicator
			int(6985)
NULL
		

Made in India. brand NAGMAN, model METCAL 40 Plus

Fast delivery to Latvia, Estonia and Lithuania. International delivery to Europe, Asia, Africa and other continents available on request
Our warehouse is located in Riga, Latvia.

price by request

Available

Available

Advantages

Official distributor in Latvia

Time-tested American, European and Asian brands of control and measurent instruments

Advantages

Special price list for organisations

We offer a significant discount for trade organisations and manufacturing companies

Advantages

Worldwide shipping on request

We take care of all logistics and customs clearance. Fast delivery to Baltic countries

Advantages

After-sales service and support

Warranty service and repair, expert assistance in the operation of the instruments

Description of NAGMAN METCAL 40 Plus

  • Precision Dry Block Calibrator with Multihole Interchangeable Thermowells
  • Ideal for quick / fast Site Calibration of RTDs / TCs / Bi-Metal Thermometers, Temperature Switches / Indicators etc. in Industrial & Marine Application
  • Measuring Accuracy : ±0.025% of rdg.
  • High Accuracy, Resolution & Stability with Measurement Capability for mV / V / mA, Resistance TC (ISA Types : J, K, R, S, N, B, T), RTD (Pt100, Pt200, Pt500, Pt1000 : IEC Standard) Signals
  • Basic Instrument
  • Test Leads
  • Mains Cable
  • Multihole Insertion Tubes (to suit 2 x ¼“ & 1 x ½“ probes)
  • Tool for Insertion Tubes
  • Spare Fuses
  • Computer Interface & “METCAL” Calibration Software
  • Calibration Certificates are issued in Accordance with our Scope as granted by NABL per ISO/IEC 17025:2017 Standards
  • Instruction Manual
  • Carrying Case

Technical specifications of NAGMAN METCAL 40 Plus

  • Range: -25°C to 120°C (@ 21°C ±2°C Ambient)
  • Resolution: 0.01°C
  • Accuracy: ±0.2°C
  • Stability: ±0.02°C
  • Stabilisation Time: 7 Minutes
  • Temperature Readout: °C / °F Selectable
  • Axial Uniformity (40 mm): ±0.2°C
  • Radial Uniformity: ±0.02°C
  • Well Diameter: 30 mm
  • Immersion Depth: 150 mm
  • Heating Time (Amb. to 120°C): 10 Minutes approx.
  • Cooling Time (Amb. to -25°C): 30 Minutes approx.
  • Switch Test facility: Provided
  • Interchangeable Thermowells: Multihole to suit 2 x ¼” & 1 x ½” Probes. Other Sizes available as optional
  • Housing: Bench-top
  • Power Supply: 230V AC, 50 Hz, 140 W
  • Instrument Dimensions (L x D x H): 160 x 365 x 450 mm
  • Instrument Weight (approx.): 9 Kg.
  • Measurement Capability: Thermocouples ISA Types J, K, R, S, N, B, T RTD Pt100, Pt200, Pt500, Pt1000
    mA 0 to 25 mA mV 0 to 100 / 1000 mV
    Volt 0 to 10 V Resistance 0 to 500Ω / 3.5 KΩ

Additional Thermowells / Insertion Tubes & Multihole – Consult us

Measurement: Range, Resolution & Accuracy

Type Range Resolution Accuracy (% of reading) 
mV (Low) 0 to 100 0.001 ±(0.025%+2µV)
mV (High) 0 to 1000 0.01 ±(0.025%+10µV)
Volts 0 to 10 0.0001 ±(0.025%+0.08mV)
mA 0 to 25 0.0001 ±(0.025%+0.4µA)
Ohms 0 to 500 0.001 ±(0.025%+12mΩ)
KOhms 0 to 3.5 0.00001 ±(0.025%+120mΩ)
RTD Pt100 -100 to 800 0.01 ±(0.025%+0.1°C)
RTD Pt200 -100 to 600 0.01 ±(0.025%+0.05°C)
RTD Pt500 -100 to 530 0.01 ±(0.025%+0.1°C)
RTD Pt1000 -100 to 650 0.01 ±(0.025%+0.1°C)
Type Range Resolution Accuracy (% of reading) 
TC Type J -100 to 1200 0.1 ±(0.025%+0.1°C)
TC Type K -60 to 1260 0.1 ±(0.025%+0.1°C)
TC Type R 150 to 1700 0.1 ±(0.025%+0.2°C)
TC Type S 170 to 1700 0.1 ±(0.025%+0.2°C)
TC Type N 0 to 1300 0.1 ±(0.025%+0.1°C)
TC Type B 920 to 1820 0.1 ±(0.025%+0.01°C)
TC Type T -100 to 400 0.1 ±(0.025%+0.1°C)

User Guide and Instruction manual for NAGMAN METCAL 40 Plus

You can request the necessary certificates, documentation, instructions and user guide for NAGMAN METCAL 40 Plus using the special form.

You can buy NAGMAN METCAL 40 Plus in Latvia at a favourable price in «OLIL SIA» by leaving a request on the website, by phone +371 672 70580 or by email . NAGMAN METCAL 40 Plus price are subject to change depending on exchange rates. We deliver instruments by transport companies in Latvia, Estonia and Lithuania. Worldwide shipping to Europe, Asia, Africa, North and South America available on request.